Pump tubing that has lost its elasticity will not deliver the sample to the nebulizer correctly. When the pump tubing appears limp or flattened, or does not require stretching to fit over the pump rollers, it should be replaced with new pump tubing.
There are three steps to install the peristaltic pump tubing:
Figure 1. Peristaltic pump with arrow showing roller rotation, drain tubing clips not shown
1 |
Top tube retainer |
2 |
Pump rollers |
3 |
Bottom tube retainer |
4 |
Pressure bar tensioner screw |
5 |
Pressure bar tensioner |
6 |
Pressure bar |
Pump rollers rotate counterclockwise. Liquid delivery to the nebulizer (2 in Figure 2) and drain vessel (1 in Figure 2) occurs out the top end of the tubing. Sample uptake from the sample bottle or autosampler (4 in Figure 2) and liquid removal from the drain port of the spray chamber (3 in Figure 2) occurs in the lower end of the tubing.
The pump tubing is held in place by the pump tubing tags, which hook into the grooves on the tube retainers.
Figure 2. Peristaltic pump tubing liquid orientation
1 |
To the drain vessel |
2 |
To the nebulizer |
3 |
From the spray chamber drain port |
4 |
From the sample bottle or autosampler |
![]() |
The peristaltic pump has moving parts that can cause damage to hands or fingers if caught underneath the clamps during operation. Keep hands away from the peristaltic pump while it is running. |
![]() |
You must first install the spray chamber, then tubing, and then adjust the flow rate. |
![]() |
Tubing and barb sizes (and whether a barb is needed) listed below may vary depending on the tubing and spray chamber used in the experiment. See www.agilent.com for ordering information. |
1 |
PVC tubing - 1/16 in ID x 1/16 in wall (3.18 mm ID x 1.59 mm wall) |
2 |
Drain pump tubing - barb and tubing sizes depend on the application requirements. |
3 |
PVC tubing - 1/16 in ID x 1/16 in wall (3.18 mm ID x 1.59 mm wall) |
4 |
To spray chamber |
5 |
Barb - 1/16 in ID x 1/16 OD (1.6 mm to 1.6 mm) |
6 |
Barb - 1/16 in ID x 1/8 OD (1.6 mm to 3.2 mm) |
7 |
To drain vessel |
![]() |
For some samples, it may be necessary to select different pump tubes and/or pump speeds to optimize performance. If other sizes of pump tubing are used, alternative fittings may be required. Always ensure the drain tubing has a greater internal diameter than the sample tubing. |
![]() |
You must first install the nebulizer, then the tubing, and then adjust the flow rate. |
1 |
Solution pump tubing* |
2 |
Capillary tubing* |
3 |
Pump tubing tags |
4 |
To solution bottle or autosampler |
* Barb and tubing sizes depend on the application requirements.
![]() |
To reduce the dead volume and maximize sample throughput, cut the excess pump tubing about 10 mm outside the tags on both ends. |
Figure 5. Solution and drain tubing
1 |
Sample capillary tubing from peristaltic pump |
2 |
Spray chamber drain tubing to peristaltic pump. |
![]() |
To help see the flow, introduce some air bubbles into the capillary by lifting the sample tubing in and out of the sample bottle. |