Instructions are provided below on how to:
![]() |
After replacing all required components, perform the Advanced Valve System test on the Instrument > Tests window. |
Component replacement intervals vary depending upon the demands of your application.
The valve has been designed so that the consumable components within the valve (the rotor and stator), can be replaced.
Unless mentioned otherwise, the instructions below apply to all three switching valves.
![]() |
This video demonstrates how to remove and replace the AVS switching valve and stator. To replace the stator, you must replace the entire switching valve body. |
![]() ![]() |
Chemical Hazard |
To remove the valve body/stator:
To install the valve body/stator:
![]() |
The valve and pin are designed so that the valve will only insert in the correct orientation. Do not force the valve. |
![]() |
After replacing all required components, perform the Advanced Valve System test on the Instrument > Tests window. |
Over time the rotor can become worn or damaged by chemicals.
![]() ![]() |
Chemical Hazard |
![]() |
The cutouts on the driver and tabs on the rotor align so that the rotor can only go in one way. |
![]() |
After replacing all required components, perform the Advanced Valve System test on the Instrument > Tests window. |
![]() ![]() |
Chemical Hazard |
![]() ![]() |
Chemical Hazard |
![]() |
Ensure that only tubing and connectors that have an outer diameter (OD) of 1.6 mm (1/16 in.) are used or leaks may occur at the ports of the switching valve. When cutting tubing ensure that the ends are flat and do not have burrs or particulates. |
Tubing kits for the AVS 4, AVS 6, and AVS 7 come with the fittings already attached except for autosampler line. If you need to replace the fittings, always use the recommended nut and ferrule combination when replacing tubing that connects to a port on the switching valve.
The fittings supplied in the tubing kits are one piece with the nut and ferrule in one.
Trim the tubing to ensure a smooth, even burr-free surface.
Always use the recommended tubing type for your application and instrumentation. Agilent supplies tubing kits preassembled with the correct fittings. Tubing that has become kinked, clogged, or damaged must be replaced.
Select from:
![]() ![]() |
Chemical Hazard |
![]() |
Ensure that only tubing and connectors that have an outer diameter (OD) of 1.6 mm (1/16 in.) are used or leaks may occur at the ports of the switching valve. When cutting tubing ensure that the ends are flat and do not have burrs or particulates. |
![]() |
Tips More thorough mixing of the sample with diluent/IS will occur with a longer piece of tubing between the valve and the nebulizer, which ensures better replicate reproducibility, but this benefit comes at the expense of the additional time required for the sample to reach the plasma. |
Where A is the AVS valve, B is the Sample loop holder, C is the AVS pump inlet, D is the AVS pump outlet, and E is the Bubble injector.
Valve port number from image "A" above |
Tubing kit part number |
Description |
1 |
G8010-60305 |
|
2 |
G8010-60304 |
|
3 |
G8010-60303 |
|
4 |
G8010-60305 |
|
5 |
G8010-60306 |
|
6 |
G8010-60308 |
|
7 |
G8010-60307 |
![]() |
After replacing all required components, perform the Advanced Valve System test on the Instrument > Tests window. |
For AVS 7 only.
Connect the Internal Standard Connection Kit, G8010-60307. The tubing is Labeled 'Valve Port 7'.
Connect the Autosampler Connection Kit, G8010-60306.
Connect the Sample Loop, G8010-60305 for the supplied 1 mL loop. Other loop sizes are available.
Connect the Sample Inlet Connection Kit, G8010-60303. This tubing kit comes assembled with three lines connected by a 'tee' piece.
To connect the Sample Inlet Connection Kit, remove the drip tray for easier access. Reinstall it once this procedure is done.
Insert the free end of the rinse inlet solution tube into the rinse bottle.
Part number G8010-60308. The kit includes two preassembled pieces of tubing.
Part number G8010-60304. Labeled 'Valve Port 2'.
![]() |
The OneNeb nebulizer is not recommended for use with the AVS 6 or AVS 7. It may cause excessive back pressure in the sample line. If you are using a nebulizer that requires a different fitting to the one supplied with the Nebulizer Connection kit, purchase the Autosampler Connection kit, part number G8010-60306, which includes the fitting (fits 1/16 inch, 1.6 mm OD tubing) needed for the switching valve. Always ensure that the nebulizer used produces minimal sample line back pressure. Click here for more information. |
Valve port number from image above |
Tubing kit part number |
Description |
2 |
G8010-60333 |
Nebulizer Adapter Kit. This kit is used with the OneNeb (as is), or in conjunction with G8010-60336 to connect a UniFit nebulizer. |
2 |
G8010-60336 |
|
2 |
|
|
1 and 3 |
G8010-60335 |
Sample/Rinse Connection Kit
|
4 |
G8010-60334 |
![]() |
After replacing all required components, perform the Advanced Valve System test on the Instrument > Tests window. |
Follow this procedure to connect the rinse tubing. Connect the one of the tubing pieces in the Sample/Rinse Connection Kit, G8010-60335 to port 3 on the AVS 4.
Connect the Unifit Nebulizer Connection Kit, G8010-60336 to port 2 on the AVS 4.
![]() |
If you are using the OneNeb nebulizer, use the Nebulizer Adaptor Kit (G8010-60333) instead of the Unifit Nebulizer Connection Kit, G8010-60336. The flangeless fitting from the Nebulizer Adaptor Kit can be used to connect any nebulizer that uses a 1/16" OD capillary for the sample line. The tubing sleeve can be used for the connection of any nebulizer that uses a 0.042" or 1 mm OD capillary for the sample line. |
Connect the Nebulizer Adaptor Kit, G8010-60333 to port 2 on the AVS 4.
![]() |
Only use this kit if you are using the OneNeb nebulizer. Otherwise, use the Unifit Nebulizer Connection Kit, G8010-60336. |
The flangeless fitting from the Nebulizer Adaptor Kit can be used to connect any nebulizer that uses a 1/16" OD capillary for the sample line. The tubing sleeve can be used for the connection of any nebulizer that uses a 0.042" or 1 mm OD capillary for the sample line.
Always cut the capillary to appropriate length to fit between the valve and nebulizer. 100 mm is the recommended minimum.
Connect the Waste Tube Kit 4 Port, G8010-60334, to port 3 on the switching valve.
Follow this procedure to connect the sample tubing. Connect the one of the tubing pieces in the Sample/Rinse Connection Kit, G8010-60335 to port 1 on the AVS 4.