AVS 4, AVS 6 and AVS 7 Switching Valve Maintenance Checklist

The following parts of the AVS 4, AVS 6, and AVS 7 require routine maintenance. Any maintenance procedures not specifically mentioned should be carried out only by Agilent-trained, Agilent-qualified or Agilent-authorized field service engineers.

Daily

  • Clean the surface of your switching valve accessories (spills should be cleaned up immediately).
  • Inspect the switching valve tubing if present. Replace if necessary.
  • Check that the sample loop is filling completely for the first few samples in the first run of the day.
  • Check there are no blockages in the uptake line, sample probe or connectors used on the uptake line.
  • Check for blockages in the switching valve. If the loop still does not fill completely, the rotor may need to be replaced.
  • Check for increases in carryover in blanks. If carryover is increasing, the valve may require the rotor and/or stator to be cleaned or replaced.
  • Flush the sample flow path to remove acidic solutions. At the end of each period of operation, always flush through the flow paths of the switching valve accessory with deionized water or an appropriate solvent for organic applications, such as isopropyl alcohol (propan-2-ol).

See also: