The SPS 4 autosampler has a state-of-the-art autosampler design and is the latest generation of autosamplers from Agilent. The SPS 4 is a random access, single probe autosampler for the front-end automation of several of Agilent's instruments.
Where:
|
Item |
Description |
1 |
Accessory port |
DA-15 receptacle For connection of Agilent-specified accessories only |
2 |
Serial port |
DE-9 plug. RS-232 port for connection of Agilent-specified accessories |
3 |
Ethernet |
Communication cable to the PC |
4 |
USB port |
USB type B receptacle. Supports USB 2.0 Full Speed SPS 4 USB connection from the autosampler |
For detailed installation information specific to the Agilent SPS 4, see the documentation supplied with your accessory.
![]() |
Tip |
![]() |
Tip |
The software will try to determine the correct port when the "Connect to PC Port" option is enabled. If the sampler is connected it typically finds the correct port and selects it by default. If it is not able to find the correct port, follow the procedure below.
![]() |
If the SPS 4 is connected to the PC instead of the instrument, and you are using a different SPS 4 autosampler to the one originally configured, the COM port will change. You must determine the new COM port using this procedure and then select that new COM port value on the Options > General page. |
The sample probe may be displaced from the center due to the sample probe hitting a sample vial or the sides of the fixed wash reservoir. User calibration allows for minor adjustments to center the sample probe on an autosampler that is misaligned. The calibration wizard enables manual re-alignment of the sample probe using four calibration points where all corners in the movement range of the probe arm are used. At each calibration point, adjust the probe position.
![]() |
The SPS 4 autosampler must be connected directly to the PC, not the ICP-OES instrument, for the Calibration Wizard to work correctly. If it is connected to the ICP-OES, before attempting to use the Calibration Wizard, remove the SPS 4 communications USB cable from the ICP-OES and plug it into the PC. Determine the COM port by following the instructions Determining the SPS 4 COM Port. |
Before starting this procedure, always check the following:
If you are unsure of any of the above, check the Agilent SPS 4 Autosampler User's Guide for detailed instructions. If checking or performing the above does not fix the issue, perform a calibration using the Calibration Wizard.
![]() |
Close the ICP Expert application software before launching the Calibration Wizard. |
To access the Calibration Wizard:
To edit the rack configuration in the Calibration Wizard:
The following describes the calibration wizard window.
The indicators next to the calibration point buttons indicate the following status:
To perform a user calibration:
The sample racks fitted should be empty and not contain any sample vials. If a microtiter plate adapter is fitted, it should be loaded with four 96 well microtiter plates.
If you chose the wrong one, an error message appears. If the appropriate COM port number is not listed, check the autosampler power, the USB connection, and the COM port setting in the instrument control software.
![]() |
During the calibration process, it is important that you do not click the Capture button while moving to the different calibration points. If you have adjusted the position to be too far away from the theoretical XY position at any point during the calibration, an error message is displayed to instruct you to try again. |
The indicator next to the “Rear Left” button changes from orange to green.
After each position has been completed, the autosampler will automatically move to the next position.
On running the calibration wizard software, you may encounter any error listed below:
Serial Port Error
Example: Autosampler did not respond.
A serial port error occurs if the software cannot communicate with the autosampler. Check that the power is on, and no error lights are illuminated. Also confirm the USB connection and the COM port setting in the instrument control software. If the problem still remains unresolved, contact your Agilent customer support representative.
Envelope Error
Example: Probe reached the maximum Y limit.
An envelope error occurs if you try to move the sample probe outside its motion envelope limits. If the well position is outside the motion envelope, contact your Agilent customer support representative.
Capture Failure
Example: Probe is too far from starting point.
A capture failure occurs if the point you are trying to capture is too far away from the expected position. Check that the sample probe positions above the correct well position, and the sample racks you are using match the description in the Calibration Wizard. If the problem still remains unresolved, contact your Agilent customer support representative.
Autosampler Error
Example: Move Z failed with ERR(8).
An autosampler error occurs if the arm has a collision, such as if the probe hits the sample rack. If this happens, click the Reset button. The autosampler will reset, and the calibration process can be continued. If the problem still remains unresolved, contact your Agilent customer support representative.
Skipping Calibration
Example: None of the positions were adjusted. Skipping calibration.
If all four captured positions are within 0.5 mm of their original position, the existing calibration is deemed to be accurate, so the calibration does not be adjusted. In this case, any problems you are having are unlikely to be caused by alignment. Contact your Agilent customer support representative.
![]() |
For autosamplers other than the Agilent SPS 3 and SPS 4 it is not always possible for ICP Expert to detect when the autosampler has gone out of alignment. Therefore it is not always possible for a run to be stopped in this situation. |