Error Flags

Each concentration result may appear with a concentration result 'Flag'. The following table lists each possible concentration result flag and its description. The flags are listed in the order in which they are prioritized, from highest to lowest. If a sample concentration results in more than one error being flagged, the flag with the highest priority is displayed.

Error flags will be displayed on the Analysis page. However, if there is more than one flag related to a sample, only the flag of highest priority (as discussed above) will be displayed.

Flag

Description

Conc/Intensity

Measurement/Replicate

--

All replicates have been excluded

Both

Measurement

####

The signal has overranged

Both

Replicate

Error

No FACT model exists

Both

Replicate

Error

Unable to calculate internal standard ratio

Both

Measurement

Uncal

The cell is uncalibrated

Concentration

Replicate

!

The solution has been remeasured

Both

Measurement

e

Solution has been edited

Both

Measurement

x

Incomplete model

Both

Replicate

o

Concentration is over the calibration range

Concentration

Replicate

u

Concentration is under the calibration range

Concentration

Replicate

i

Internal standard not run yet

Both

Measurement

#

An overrange is present

Both

Replicate

m

Unable to model the measurement

Both

Replicate

W

QC spike failure

Concentration

Measurement

*

Duplicate sample failure

Concentration

Measurement

L

LCS failure

Concentration

Measurement

Q

CCV, ICV failure

Concentration

Measurement

N

Matrix spike failure

Concentration

Measurement

Z

ICB, CCB failure

Concentration

Measurement

K

Interferent Check Standard A

Concentration

Measurement

G

Interferent Check Standard AB

Concentration

Measurement

E

Serial Dilution

Concentration

Measurement

R

CRDL limit failure

Concentration

Measurement

Y

Prep blank failure

Concentration

Measurement

S

QC setup failure

Concentration

Measurement

J

LSpike failure

Concentration

Measurement

A1, A2...

All Sample QC failure

Concentration

Measurement

P

Component - Inhalational, Oral or Parenteral

Concentration

Measurement

CT

Control Threshold - Inhalational, Oral or Parenteral

Concentration

Measurement

P

Product - Inhalational, Oral or Parenteral

Concentration

Measurement

b

Line has a related interferent that is above the extended calibration range (IEC correction still applied)

Concentration

Replicate

p

Internal modeling only

Concentration

Replicate

There are a number of default QC failure flag names, which are listed on the QC Test types and Failure flags page.