Click here for information on how to setup and run methods.
Click here for a software overview that includes information about the Analysis page.
![]() |
All features on the Analysis page are available in both the ICP Expert and ICP Applet view. |
The Analysis page of the worksheet is where you can select samples for analysis, and collect, view and edit analytical results.
![]() |
Where: 1. Results table 2. Spectrum display 3. Replicate and Calibration data, Interferences, Outlier setup, USP validation (not shown on this image), and the Operations Log and 4. Calibration graph and Trend Analysis |
Select the column or columns to change the units for the selected columns, edit the number of significant figures and decimal places displayed, select whether a column is visible and which results flags are displayed in the results table.
To select multiple columns:
Add/Remove/Edit Custom Columns
![]() |
Always add a space after entering a number. |
To delete the custom column from the results table right-click the custom column and then select Remove Custom Column.
All custom column calculations convert all column values to "g/mL" before the equation is evaluated. For example, results in ppm/mg/L etc will be divided by 1000000, % will be divided by 100, etc before evaluation of the equation. The units selected in the Column Properties can then be used to convert the result to an appropriate unit (as defined in the "Units" definitions under "Options").
This will appear only if Enable MultiCal view and MultiCal summary concentration are selected on the Standards page.
To select the column wavelengths and calculation method:
![]() |
When using 21 CFR Part 11 software, Edit MultiCal Column is enabled only if the user has the Results Editing privilege. Any change when using these features is added to the CFR Audit Trail. |
Deselect this option to hide the IEC block from the sequence, freeing up sample tubes in the standards rack the autosampler.
![]() |
This option is only visible/available for worksheets with IEC enabled on the Elements page, defined interferent wavelengths and existing IEC calibration data. |
Click the down arrow or button to choose whether to display data as concentration, intensity or unadjusted concentration results.
Select to track the analysis as it progresses. Deselect to view selected data.
Click the down arrow to choose whether to display results in the default order, alphabetically, by wavelength or by type.
Use to show or hide specific columns. Select a column, or press SHIFT or CTRL and left-click to select multiple columns, to hide. Available selections are dependent on how the worksheet is setup (whether there are Internal Standards, a MultiCal summary, etc.)
See the description above for more information about column properties.
Click the arrow next to the Measurement Selection icon to select between the following measurement display options.
Default has the following behavior.
Summary has the following behavior.
Displays the last measurement taken.
Click this to remove all results from the worksheet. You can also press CTRL+DELETE to delete results. Once the results are deleted, you can not retrieve them.
When using the 21 CFR Part 11 software, you cannot delete results.
Click the appropriate icon to display (or hide) the Results table, Spectrum Display, Replicates/Calibration Data table and the Calibration Graph.
Right-click the results cell to access more features including Add/Remove Duplicate, Export Selected Solutions, Measure and Set Sample Units.
Add/Remove Duplicate
Adds a duplicate column of the selected element or removes the selected duplicate column.
Export Selected Solutions
Exports the selected solutions into a .csv file format. Microsoft Excel format is also available in the ICP Expert Pro feature pack.
![]() |
When using 21 CFR Part 11 software, Adding, Removing, or Editing Custom Columns is enabled only if the user has the Results Editing privilege. Any change when using these features is added to the CFR Audit Trail. |
See the Autodilution page for information about this feature.
See Spectrum display for information on the spectrum.
This is the default view of the table in the lower left corner of the Analysis page and is accessed by clicking the Replicate Data icon. Error or solution flags will appear next to this table.
All measurements below this header are statistics relating to the solution concentrations.
All measurements below this header are statistics relating to the solution intensities.
Shows the average of both the concentration and the intensity for all selected replicates.
Displays the standard deviation of all selected replicates for both the concentration and intensity.
Displays the relative standard deviation of all the selected replicates for both the concentration and intensity. This RSD value is calculated on the intensities. The percentage Relative Standard Deviation (% RSD) is a measure of the precision of the measurement (or how reproducible the result is) when multiple replicate readings are made on a given sample. The precision is dependent on the intensity of the signal, the care with which samples and standards are prepared, and the stability of the optical system.
Select which replicates are to be included in the statistical calculations and the analysis worksheet table. This is used to edit out, or select the replicates of a selected element (the result cell is selected from the worksheet display). Once a result cell is selected, its replicates and the mean results are displayed in the replicates panel.
Shows the number of replicates.
The concentration column displays either the set concentration or the calculated concentration for each replicate. If you are looking at a standard, you will see the concentration shown as the defined concentration. For all samples and QC you will see the calculated concentration.
When a standard is selected in the analysis table while Replicate Data is selected, Concentration changes to Method Concentration. The table then displays the user assigned concentration or each standard used in the calibration curve.
Displays the intensity of each replicate.
When a standard is selected in the analysis table while Replicate Data is selected, this column appears and displays the actual concentration according to the calibration equation for each standard in the calibration curve.
Shows the statistics related to the calibration fit of the element selected. It is accessed by clicking the Calibration Data icon.
Shows all the standards included in developing the calibration curve for this element line.
Indicates which measurement results are displayed. If the standards were autodiluted, selecting the diluted standard in the analysis table would change the Measurement information displayed.
Shows the average intensity of each standard used in creating the calibration curve.
Shows the user assigned concentration or each standard used in the calibration curve.
Shows the actual concentration according to the calibration equation for each standard in the calibration curve.
The % difference between the method concentration and the actual concentration for each standard used in the calibration curve.
These are the same Interferences table and graph
shown on the Elements page.
Shows the operations that have taken place in the ICP Expert software. It is accessed by clicking the Show operations log icon.
This is accessed by clicking the Outlier Setup icon . Enable and set the Pass Value for 6 different Outlier Flag Rules. When the Pass Value is not met, the affected results in the Results table are color coded for easy identification. Choose Display flagged only to show only the results that did not meet the selected rule.
Example of two rules selected and applied to the results
Available only in ICP Expert Pro or ICP Expert with 21 CFR 11 enabled. Enable USP <232/233> and ICH Q3D expanded QC on the File > Options menu and Configuration tab. Click the USP Validation icon to access this section.
It includes a drop-down list for all USP method validation tests. For more information, see the USP <232/233> and ICH Q3D page.
The calibration graph displays the calibration graph for the element or wavelength selected in the element table. Right-click to Export the graph or apply graph smoothing.
Displays the argon intensity and nebulizer pressure variations for each sample analyzed. It is accessed by clicking on the Trend Analysis icon on the Calibration graph in the lower right corner of the window.
This is useful for troubleshooting and to identify changes over time that may indicate maintenance is required.
This feature is only available when using ICP Expert Pro.
It is accessed by clicking on the Internal Standard Ratios icon on the Calibration graph in the lower right corner of the window.
Appears only when at least one wavelength in the worksheet is an internal standard. This is useful to easily identify changes in internal standard ratio over the course of a worksheet run. The chart plots the ratios of all of the wavelengths which have been assigned the 'Internal Standard' type on the Elements page on a per-solution basis. Hide and display the data for each internal standard wavelength by selecting or deselecting the wavelength check box.