The Calibration tab is accessed by clicking 'Calibration' in the navigation panel on the left of the Instrument window.
![]() |
A detector calibration can only be performed with the plasma off and the chiller will need to be on to cool the camera to -40 °C. The Instrument Calibrate button will be unavailable until the detector calibration is done. |
The detector calibration performs a dark current calibration to correct for detector and electronic background signals.
To perform a detector calibration, once the camera has reached the correct temperature, click Calibrate in the Detector section.
The following information explains how to perform an instrument calibration. The calibration consists of a wavelength calibration. To perform the instrument calibration, you will need to use a pre-made calibration solution or make up a calibration solution. To do this see Preparing the instrument calibration solution.
An instrument calibration needs to be performed the first time the instrument is run, when new software is loaded, when drift correction lines are added or removed, and when the Polychromator purge gas is changed. The gases that can be used for purge are either Ar or N2.
It is recommended that an instrument calibration be performed monthly thereafter.
![]() |
An instrument calibration can only be performed if a valid detector (dark current) calibration has been performed on the instrument. |
Before starting the system, carefully read the Safety practices and hazards section in your ICP-OES User's Guide and ensure that the laboratory is set up according to the specifications listed in the Agilent ICP-OES Site Preparation Guide. Check that the snout, cone, torch and all tubing on the spray chamber, nebulizer and peristaltic pump are correctly connected.
To perform an instrument calibration:
![]() |
Ensure a standard glass concentric nebulizer, a double pass spray chamber and the appropriate torch are installed. Use manual sampling. Use 'white/white' peristaltic pump solution tubing and 'blue/blue' tubing for the drain. The tubing tab color denotes the tubing size. |
![]() |
The plasma will take approximately 60-80 seconds to ignite. If it fails to ignite, see Plasma not igniting for further information. |
![]() |
Tip For optimum analytical performance, it is recommended that the plasma is turned on from standby mode at least 20 minutes prior to analysis to allow it to stabilize. If the instrument has been turned off for an extended period of time, a longer warm up may be necessary. |
Leave the polychromator boost purging for at least 20 minutes prior to performing the instrument calibration.
If you are not going to be analyzing UV wavelengths below 189 nm, you can choose not to have the polychromator boost and snout purge on. However, wavelengths below 189 nm will have a red cross next to them on the results table.
*It is highly recommended that the Polychromator Boost purge and Snout (when using the radial viewing mode) purge be on when performing an instrument calibration. If you do not intend to measure wavelengths under 189 nm, turn off the Polychromator Boost and Snout purge after the calibration is finished.
After a short while, an indication of the success or failure of the calibration will appear, as well as an indication of the wavelength offsets.
The results of the wavelength calibration are stored as text files and are saved to text files entitled 'Calibrationdetaillog.txt' and 'Calibrationsummarylog.txt' located in the following directory:
C:\ProgramData\Agilent\ICPExpert\Log\Instrument serial number\Result.txt
![]() |
Tip If the calibration failure is with the low UV wavelengths such as Al it is most likely that a longer polychromator (optics) purge time is required or the Polychromator boost is off and/or the snout purge is off. If the failures are more wide spread, prepare a new wavelength calibration solution and try again. You can also examine the Calibrationdetaillog.txt or Calibrationsummarylog.txt file and determine which elements failed the calibration. |
After a short while, an indication of the success or failure of the calibration will appear, as well as an indication of the wavelength offsets.
The results of the wavelength calibration are stored as text files and are saved to text files entitled 'Calibrationdetaillog.txt' and 'Calibrationsummarylog.txt' located in the following directory:
C:\ProgramData\Agilent\ICPExpert\Log\Instrument serial number\Result.txt
![]() |
Tip If the calibration failure is with the low UV wavelengths such as Al it is most likely that a longer polychromator (optics) purge time is required or the Polychromator boost is off and/or the snout purge is off. If the failures are more wide spread, prepare a new wavelength calibration solution and try again. You can also examine the Calibrationdetaillog.txt or Calibrationsummarylog.txt file and determine which elements failed the calibration. |
After a short while, an indication of the success or failure of the calibration will appear, as well as an indication of the wavelength offsets.
The results of the wavelength calibration are stored as text files and are saved to text files entitled 'Calibrationdetaillog.txt' and 'Calibrationsummarylog.txt' located in the following directory:
C:\ProgramData\Agilent\ICPExpert\Log\Instrument serial number\Result.txt
![]() |
Tip If the calibration failure is with the low UV wavelengths such as Al it is most likely that a longer polychromator (optics) purge time is required or the Polychromator boost is off and/or the snout purge is off. If the failures are more wide spread, prepare a new wavelength calibration solution and try again. You can also examine the Calibrationdetaillog.txt or Calibrationsummarylog.txt file and determine which elements failed the calibration. |
After a short while, an indication of the success or failure of the calibration will appear, as well as an indication of the wavelength offsets.
The results of the wavelength calibration are stored as text files and are saved to text files entitled 'Calibrationdetaillog.txt' and 'Calibrationsummarylog.txt' located in the following directory:
C:\ProgramData\Agilent\ICPExpert\Log\Instrument serial number\Result.txt
![]() |
Tip If the calibration failure is with the low UV wavelengths such as Al it is most likely that a longer polychromator (optics) purge time is required or the Polychromator boost is off and/or the snout purge is off. If the failures are more wide spread, prepare a new wavelength calibration solution and try again. You can also examine the Calibrationdetaillog.txt or Calibrationsummarylog.txt file and determine which elements failed the calibration. |