You can view the status of several instrument components in the Instrument Window Status tab. Click any flashing component on the instrument image to go to troubleshooting for that component.
Follow the procedure below to see details about the instrument viewing mode and the instrument and component serial numbers, hardware, and firmware versions. These numbers can only be edited by your Agilent Field Representative.
Manually control several instrument components in the Instrument Window Status tab.
These are manual control of flow rates only and cannot be adjusted while a worksheet is running.
Set desired and view actual gas flow rates:
How do I adjust these settings in my worksheet?
Oxygen injection is only available when using ICP Expert Pro.
![]() |
Use premixed 80%/20% Argon/Oxygen only. |
The Cone purge is automatically turned on when the plasma is turned on.
These are manual controls only and cannot be adjusted while a worksheet is running.
This can only be set when the plasma is on. This is a manual control only and cannot be adjusted while a worksheet is running.
This is a manual control only and cannot be adjusted while a worksheet is running.
This can only be set when the plasma is on.
Select the beam selector position and enter the height on the Conditions page in a worksheet.
This is useful to monitor to ensure that you don't start an analysis before the polychromator has reached 35 °C. This is display only and is not adjustable.
Use this to change how fast to operate the peristaltic pump. This is a manual control only and cannot be adjusted while a worksheet is running.
How do I set the Fast Pump speed for my analysis?
Use this to check that the valve switches, and whether the uptake pump and bubble injector work.
If you have replaced the valve or rotor, click Reset to ensure the valve is in the correct position.
Use this to change the IsoMist temperature. This is a manual control only and cannot be adjusted while a worksheet is running.