![]() |
To access rate driven QC, you must have the ICP Expert Pro package installed. Otherwise, all QC solutions are non-rate driven. |
This page provides information on the following:
A rate driven QC block is created on the Sequence page by selecting, dragging and then dropping the required QC solutions after samples or other rate driven QC blocks into the Sequence table on the Sequence page or by selecting and then editing the 'Number of samples between QC tests' in the Rate QC Parameters dialog box.
The table below indicates default behaviors when dragging a QC test into the sample list.
QC test |
Drag over |
Default preference |
Rack allocation |
Over calibration |
Insert after every calibration |
Standards |
|
Over sample |
Number of samples between QC tests |
||
Over QC |
Number of samples between QC tests |
||
Over calibration |
Insert after every calibration |
Standards |
|
Over sample |
Number of samples between QC tests |
||
Over QC |
Number of samples between QC tests |
||
Over calibration |
Insert after every calibration |
Standards |
|
Over sample |
Number of samples between QC tests |
||
Over QC |
Number of samples between QC tests |
||
Over calibration |
Insert after every calibration |
Standards |
|
Over sample |
Number of samples between QC tests |
||
Over QC |
Number of samples between QC tests |
||
Over calibration |
Non-rate driven |
Sample |
|
Over sample |
Non-rate driven |
||
Over QC |
Non-rate driven |
||
Over calibration |
Not droppable |
Sample |
|
Over sample |
Non-rate driven |
||
Over sample related QCs |
Non-rate driven |
||
Other QCs |
Not droppable |
||
Over calibration |
Not droppable |
Sample |
|
Over sample |
Non-rate driven |
||
Over sample related QCs |
Non-rate driven |
||
Other QCs |
Not droppable |
||
Over calibration |
Non-rate driven |
Sample |
|
Over sample |
Non-rate driven |
||
Over QC |
Non-rate driven |
||
Over calibration |
Non-rate driven |
Sample |
|
Over sample |
Non-rate driven |
||
Over QC |
Non-rate driven |
||
Over calibration |
Not droppable |
Sample |
|
Over sample |
Non-rate driven |
||
Over sample related QCs |
Non-rate driven |
||
Other QCs |
Not droppable |
||
Over calibration |
Not droppable |
Sample |
|
Over sample |
Non-rate driven |
||
Over sample related QCs |
Non-rate driven |
||
Other QCs |
Not droppable |
||
Over calibration |
Non-rate driven |
Sample |
|
Over sample |
Non-rate driven |
||
Over QC |
Non-rate driven |
||
Over calibration |
Non-rate driven |
Sample |
|
Over sample |
Non-rate driven |
||
Over QC |
Non-rate driven |
The ICP Expert software provides templates to help you run analyses based on the requirements of the United States Environmental Protection Authority (US EPA) Method 200.7 and Contract Laboratory Program (CLP) Inorganic Superfund Methods (ISM02.1) protocols.
Non-rate driven solutions are defined as those that have been entered at a fixed location. They will only be run if they are selected on the Analysis page. In an analysis the QC solutions will always be run directly after the Solution Label to which they are attached.
Non-rate driven QC solutions and blocks include:
*These can be edited to become rate driven solutions in the Rate QC Parameters dialog box.
In contrast to non-rate driven solutions, rate driven solutions are defined as those that are automatically entered at a location, which is dependent upon the overall analysis sequence chosen. For example, ‘Recalibrate every 5 samples’ is a rate driven option that means the standard/s will be automatically inserted after every fifth sample in the sequence. The sequencing of these solutions is set and they will be run as required. Therefore the samples may not necessarily occur at the location in which they appear in the sequence.
When the QC solutions listed above are selected, dragged and then dropped into the samples table on the Sequence page, they are automatically considered a rate driven QC block. The Rate QC Parameters dialog box opens so that these solutions can be configured. By default, ICP Expert will work out the number of samples the block has preceding it, and use it as the rate (where the same QC block will appear throughout the rest of the sequence).
Rate driven QC solutions and blocks include:
**These can be edited to become non-rate driven solutions in the Rate QC Parameters dialog box.