![]() |
The QC tab will only be displayed if Enable QC is selected in the Configuration page. The tests are selected from the QC page. |
Select one of the QC solutions from the QC Type drop-down menu on the QC tab in the worksheet. The QC solution types can split into four types of QC solutions categories. Each of the QC solution categories has a corresponding table that allows the user to insert the QC pass criteria.
It is provided for backwards compatibility with ILMO5.x and earlier.
Laboratory Control Sample, Continuing Calibration Verification, Interference Check Solution A, Interference Check Solution AB, Contract Required Quantitation Limit Standard and Initial Calibration Verification are all check standards type solutions and share the same categories in the table that appears when one of these solution types is selected.
The following Check Standard tests are available:
The Check Standard tests form part of the suite of QC Tests available on ICP Expert.
If you enable the CCV test, the QC grid will display the following by default:
Check Box |
Select which elements from the analysis are going to be governed by the QC pass criteria. All elements are automatically selected when the test is enabled. You can select any you want, but if you have not selected an element, it will not be QC tested. |
Defined Concentration |
Enter the known concentration of the CCV. The units used are the same as those defined in the Concentrations table on the Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
The percentage recovery is calculated for the CCV as follows:
(MeasuredConc / DefinedConc) * 100
where: |
MeasuredConc is the measured concentration |
The selected error action is applied when the found % recovery is greater than the high limit or less than the low limit. The default failure flag is Q.
Continuing Calibration Verifications can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
A type of QC Standard, an ICV is a solution of known concentration used to verify the accuracy of a calibration. The ICV should be performed immediately after every full calibration.
If you enable the ICV test, the QC grid will display the following by default:
Check Box |
Select which elements from the analysis are going to be governed by the QC pass criteria. All elements are automatically selected when the test is enabled. You can select any you want, but if you have not selected an element, it will not be QC tested. |
Defined Concentration |
Enter the known concentration of the ICV. The units used are the same as those defined in the Concentrations table on the Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
The percentage recovery is calculated for the ICV as follows:
(MeasuredConc / DefinedConc) * 100
where: |
MeasuredConc is the measured concentration |
The selected error action is applied when the found % recovery is greater than the high limit or less than the low limit. The default failure flag is Q.
Initial Calibration Verifications can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
An Interference check (ICS) is used to verify that any interferent corrections performed are valid. The check requires that ICSA and ICSAB solutions are placed consecutively in the sequence.
![]() |
Interference Check Solutions are only required if you are performing Inter-Element Corrections. Click here for information on the Inter-element Correction page. |
Standard A contains the interferent element(s) in known concentration(s).
Interference Check Solution A does not use the same Limits table as the other standards. Instead, it uses the CRQL value in the 'Pass test' equation. Click here for more information about CRQL values.
If you enable the ICSA test, the table will display the following by default:
Check Box |
Select this check box to include an element in the QC test. All elements are automatically selected when the test is enabled. |
The test passes if the measured concentration is less than the CRQL.
Define the limit for the ICSA in the Detection Limits table (CRQL or MDL). If you use MDL for your limit, you must change the equation to reflect this.
The default failure flag for ICSA is K.
Interference Check Solutions can be included in an analysis by selecting them on the QC page. The action taken will depend upon the Error Action selection made on this page.
Standard AB contains the interference element(s) in Interference Check Solution A in the same concentration plus a known concentration of analyte element(s).
If you enable the ICSAB test, the QC grid will display the following by default:
Check Box |
Select which elements from the analysis are going to be governed by the QC pass criteria. All elements are automatically selected when the test is enabled. You can select any you want, but if you have not selected an element, it will not be QC tested. |
Defined Concentration |
Enter the concentration of the ICSAB. The units used are the same as those defined in the IEC or Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
The default failure flag for ICSAB is G.
The check for when the Interference Check Solutions (A and AB) are analyzed together:
MeasuredConc / DefinedConc * 100
where: |
MeasuredConc is the measured concentration |
Interference Check Solutions can be included in an analysis by selecting them on the QC page. The action taken will depend upon the Error Action selection made on this page.
The Laboratory Control Sample (LCS) is an independent check sample very similar to the QC Standard concept. The composition is known and the % recovery can be calculated and tested against the specified upper and lower recovery limits.
If you enable the LCS test, the QC grid will display the following by default:
Check Box |
Select which elements from the analysis are going to be governed by the QC pass criteria. All elements are automatically selected when the test is enabled. You can select any you want, but if you have not selected an element, it will not be QC tested. |
Defined Concentration |
Enter the known concentration of the LCS. The units used are the same as those defined in the Concentrations table on the Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
The US EPA requires a Laboratory Control Sample (LCS) to be analyzed for each group of samples in a Sample delivery group or for each group of samples digested, whichever is more frequent.
The percentage recovery is calculated for the LCS as follows:
(MeasuredConc / DefinedConc) * 100
where: |
MeasuredConc is the measured concentration |
The selected error action is applied when the found concentration of the LCS % is greater than the high limit or less than the low limit. The default failure flag is L .
Laboratory Control Samples can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
![]() |
This solution only applies to CLP ILM protocols and is obsolete for ISM protocols (although some labs may still be complying with older protocols where this was used). This solution is for compatibility with earlier ILM specifications and the term CRQL is equivalent to CRDL for ILMO4.x and earlier. It is provided for backwards compatibility with ILMO5.x and earlier. |
If you enable the CRS test, the QC grid will display the following by default:
Check Box |
Select this check box to include an element in the QC test. All elements are automatically selected when the test is enabled. |
Defined Concentration |
Enter the known concentration of the CRS. The units used are the same as those defined in the Concentrations table on the Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
The percentage recovery is calculated for the CRS as follows:
(MeasuredConc / DefinedConc) * 100
where: |
MeasuredConc is the measured concentration |
The selected error action is applied when the found concentration of the CRS % is greater than the high limit or less than the low limit. The default failure flag is R.
CRS's can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
Continuing Calibration Blank, Preparation Blank, and Initial Calibration Blank are all Check Blanks type solutions and share the same categories in the table that appears when one of these solution types is selected. Check Blanks are used to periodically monitor the Calibration Zero solution which in turn monitors the contamination levels. The Check Blank must be analyzed after every Check Standard analysis.
A QC blank is used to:
The following Check Blanks tests are available:
The Preparation Blank is a volume of distilled, de-ionized water that is processed through the sample preparation procedures. The Preparation Blank result therefore gives an indication of any contamination picked up during the sample preparation process.
If you enable the PBLK test, the QC grid will display the following:
Check Box |
Select to include an element in the QC test. All elements are automatically selected when the test is enabled. |
The test passes if the measured concentration is less than the CRQL.
Define the limit for the PBLK in the Detection Limits table (CRQL or MDL). If you use MDL for your limit, you must change the equation to reflect this.
The US EPA requires one PBLK to be analyzed with every sample delivery group or with each batch of samples digested, whichever is more frequent. If the absolute PBLK concentration is less than the CRQL or MDL limit, no action is required.
If any analyte concentration is above the DL, the lowest concentration of the analyte in the associated samples must be 10 times the PBLK concentration or else those samples which are less than 10 times the PBLK concentration must be re-digested and re-analyzed for that analyte.
The selected error action is applied when the found concentration of the PBLK is found to be greater than the specified limit. The default failure flag is Z.
Preparation Blanks can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
An ICB is a check blank that is analyzed with an Initial Check Standard. A CCB is a check blank that is analyzed with a Continuing Check Standard.
![]() |
The following information applies to both Initial and Continuing Calibration blanks. |
If you enable an ICB or CCB test, the QC grid will display the following:
Check Box |
Select this check box to include an element in the QC test. All elements are automatically selected when the test is enabled. |
The test passes if the measured concentration is less than the CRQL. Define the limit for the ICB or CCB in the Detection Limits table (MDL or CRQL). If you use MDL for your limit, you must change the equation to reflect this.
If the concentration of the ICB or CCB is found to be greater than the CRQL or MDL limit, the following procedure should occur:
![]() |
The requirements below can be met by setting the Error Action option in the QC page to Recalibrate and repeat. |
The selected error action is applied when the found concentration of the ICB or CCB is found to be greater than the specified limit. The default failure flag is Z.
Initial Calibration Blanks and Continuing Calibration Blanks can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
Paired sample tests compare the results of a pair of samples. The Duplicate and Serial Dilution solutions are Paired Samples type solutions. These solutions are the only solutions that need to be placed immediately following the solution that it is a duplicate of. This has it's own unique table that appears when this solution type is selected.
A Duplicate test involves splitting a sample into two and processing each half of the sample through the same sample preparation procedure, in order to determine the precision of the method. The duplicate will always follow the original sample.
If you enable the DUP test, the table will display the following by default:
Check Box |
Select this check box to include an element in the QC test. All elements are automatically selected when the test is enabled. |
Difference |
Enables you to set the allowable % difference limit between the sample and its corresponding duplicate. The US EPA defines the difference for Dup tests as being 20%, and this value is the default setting for this test. |
The percent difference is calculated and is then compared to the specified limit:
abs(MeasuredConc(PreviousSample) - MeasuredConc) / ((MeasuredConc(PreviousSample) + MeasuredConc) / 2) * 100
where: |
MeasuredConc(PreviousSample) is the first sample value |
The check is deemed to fail when the percentage difference exceeds the limit specified. The default failure flag is *.
Duplicates can be included in an analysis by selecting them on the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
![]() |
By default, Duplicates are counted as samples. |
A serial dilution consists of a comparison of the results of a sample and another aliquot diluted by a known factor. The pair of solutions must be placed consecutively in the sequence, with the undiluted sample being followed by the diluted sample.
If you enable the SER test, the QC grid will display the following by default:
Selected |
Select this check box to include an element in the QC test. All elements are automatically selected when the test is enabled. |
% Difference |
Enables you to set the allowable % difference limit between the two serially diluted samples. The US EPA defines the difference for SER tests as being 10%. |
When performing a Serial Dilution check, ICP Expert will do the following:
The percentage difference of the diluted sample to the preceding sample is calculated as follows:
abs(MeasuredConc - MeasuredConc(PreviousSample)) / MeasuredConc(PreviousSample) * 100
and is then compared to the limit specified.
The check is deemed to fail when the percentage difference exceeds the limit specified. The test is deemed to pass if the %D is < the limit or the previous sample concentration is < 50xMDL. If the sample concentration is greater than 50xMDL then the %D must be less than the limit to pass, if it is less than 50xMDL then the %D does not need to be less than the limit to pass. The default failure flag is E.
Serial dilutions can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made on the QC page.
Matrix Spike, Matrix Spike Duplicate and QC Spike are Spike solutions that share the same categories in the table that appears when one of these solution types is selected.
The following Spike tests are available:
A QC Spike is a Post-digest spike which is used to determine whether any significant interferences are present, thereby requiring the use of Standard Additions.
A known amount of analyte is added to the sample after the sample has been digested. The % Recovery is calculated and tested against the specified upper and lower recovery limits.
The QC Spike sample is placed immediately after the matching sample or the matrix spike of the same sample.
If you enable the Spike test, the QC grid will display the following by default:
Check Box |
Select which elements from the analysis are going to be governed by the QC pass criteria. You can select any you want, but if you have not selected an element, it will not be QC tested. |
Defined Concentration |
Enter the concentration of the relevant spike. The units used are the same as those defined in the Concentrations table on the Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
In a QC spike check, the concentration values for both the QC spike sample and its preceding sample are determined.
(MeasuredConc - MeasuredConc(PreviousSample)) / DefinedConc * 100
where: |
MeasuredConc is the spiked sample result |
Ideally, where the sample matrix is equivalent in the two samples and in the calibration standards, the percentage recovery should equal 100.
The test is deemed to fail if the percentage is found to be greater than the upper limit or less than the lower limit. The default failure flag is W.
QC Spikes can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made in the QC page.
A Matrix Spike is a Pre-digest spike. An aliquot of the sample spiked with known quantities of specific compounds is subjected to the entire analytical procedure, in order to indicate the appropriateness of the method for the matrix by measuring recovery. The % Recovery of the analyte indicates the effects of the preparation procedures on the recovery of the analyte.
If you enable the MSPK test, the QC grid will display the following by default:
Check Box |
Select which elements from the analysis are going to be governed by the QC pass criteria. You can select any you want, but if you have not selected an element, it will not be QC tested. |
Defined Concentration |
Enter the concentration of the relevant spike. The units used are the same as those defined in the Concentrations table on the Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
In a Matrix spike check, the concentration values for both the QC spike sample and its preceding sample are determined.
(MeasuredConc - MeasuredConc(PreviousSample)) / DefinedConc * 100
where: |
MeasuredConc is the spiked sample result |
Ideally, where the sample matrix is equivalent in the two samples and in the calibration standards, the percentage recovery should equal 100.
The test is deemed to fail if the percentage is found to be greater than the upper limit or less than the lower limit. The default failure flag is N.
Matrix Spikes can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made in the QC page.
If you enable the MSPK test, the QC grid will display the following by default:
Element |
Select which elements from the analysis are going to be governed by the QC pass criteria. You can select any you want, but if you have not selected an element, it will not be QC tested. |
Defined Concentration |
Enter the concentration of the relevant spike. The units used are the same as those defined in the Concentrations table on the Standards page. |
Lower Limit |
This is the lower acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Lower Limit is set to 90% and your defined concentration is 1.0, then a read back greater than or equal to 0.9 (while still satisfying the Upper Limit requirements) would result in a QC test Pass. |
Upper Limit |
This is the upper acceptable limit for the QC read back listed as a percentage of the actual concentration entered in the defined concentration cell. For example, if the Upper Limit is set to 110% and your defined concentration is 1.0, then a read back less than or equal to 1.1, (while still satisfying the Lower Limit requirements) would result in a QC test Pass. |
% Difference |
Enables you to set the allowable % difference limit between the sample and its corresponding duplicate. Default value is 20% difference. |
The default failure flag is M .
The equation for MSD is:
Matrix Spike
(MeasuredConc - MeasuredConc(PreviousSample)) / DefinedConc * 100
Duplicate
abs(MeasuredConc(MSPK) - MeasuredConc) / ((MeasuredConc(MSPK) + MeasuredConc) / 2) * 100
where: |
MeasuredConc is the spiked sample MSD result |
The test is deemed to fail if the percentage is found to be greater than the upper limit or less than the lower limit or if the % Difference is greater than the specified limit.
Matrix Spike Duplicates can be included in an analysis by selecting them from the Sequence page. The action taken will depend upon the Error Action selection made in the QC page.
Elements |
Lists the names of the elements included in the Method. |
Units |
Units reflect the selected units for this element from the Standards page. |
Estimated DL |
The estimated detection limit (EDL) value given in the software is determined based on the calibration curve for each wavelength. It’s calculated by taking 3x the calibration blank SD and converting it to a concentration. Estimated DL is only meant to be indicative which allows you to use some of the other functionality in the software that rely on MDL values (like the outlier conditional formatting rules and some QC equations) without needing to determine an MDL. The EDL value will automatically populate into the MDL column if the MDL column is blank. |
MDL |
Enables you to enter the Method Detection Limit (MDL). MDLs are required for the Serial Dilution QC test, which uses the MDL value as part of the test definition. The MDL is defined as the minimum concentration that can be measured and reported with 99% confidence that the analyte concentration is greater than zero and is determined from the analysis of a sample in a given matrix containing the analyte.1 |
CRQL |
Enables you to enter the Contract Required Quantitation Limit (CRQL). The CRQL is defined as "the minimum level of quantitation acceptable under the contract statement of work (SOW)".2 CRQLs are required for the following QC tests, which use the CRQL value as part of their test definition:
The CRQL entered for an element will be used in all relevant QC tests selected for the current method. Quantitation Limits are reported in the same units as those of their related element. The units for these elements are set in the Standard Concentrations table on the Standards page. |
References: