![]() |
ADS 2 system control is only available with ICP Expert with the Pro feature pack installed. |
The ADS 2 offers automated control of known and unknown dilution needs.
Additional information can be found on the following pages:
Solution flows are handled by the autosampler, AVS 6/7, and the ADS 2 to help automate your analysis. This section describes how the solutions are delivered during all phases of an analysis.
The video below describes the solution flow paths for both a non-diluted and diluted sample.
![]() |
This video demonstrates the solution flow paths when using an ADS 2 and AVS 6/7 with Autodilution enabled during an analysis. |
Where:
1. Nebulizer | 2. AVS pump | 3. AVS 6/7 switching valve | 4. AVS sample loop | 5. Peristaltic pump on the ICP-OES |
6. Autosampler | 7. Waste | 8. Valve C on the ADS 2 | 9. Valve A on the ADS 2 | 10. Valve B on the ADS 2 |
11. Dilution loop | 12. Diluent syringe | 13. Carrier syringe |
Valve A positions | Valve B positions | ||
Dilute | Bypass | Bypass | Load |
![]() |
![]() |
This phase is not shown the video linked above.
Initialization is performed in four steps and runs through the syringe, valve positions and flow paths.
Step 1 - flushes the syringes and cleans the tubing
The ADS 2 Valve A is in the dilute position. Valve B is in the load position. The AVS valve is in the uptake position.
Step 2 - reload syringes and cleans the tubing
The ADS 2 Valve A is in the bypass position. Valve B is in the load position. The AVS valve is in the uptake position.
Step 3 - prime syringes
Valve A is in the dilute position. Valve B is in the load position. The AVS valve is in the inject position.
Step 4 - clean AVS sample loop and nebulizer
Both ADS 2 valves are in the bypass position. The AVS valve is in the inject position.