If the problem persists after trying the following, contact your Agilent field service engineer or representative.
Error Message |
Instrument action and status LED |
Explanation |
Remedy |
Plasma extinguished due to chassis alarms (State %u) |
N/A |
The plasma was extinguished due to an alarm relating to cooling water. The cooling water supply is not switched on. |
Consult the documentation supplied with the chiller for information. |
The water supply delivery pressure is not within the required range. |
Consult the documentation supplied with the chiller for information. |
||
The inlet and outlet water hoses are connected in reverse. |
|||
There is a problem with the inlet water temperature. |
Consult the documentation supplied with the chiller for information. |
||
The water inlet filter on the instrument is blocked. |
|||
Plasma extinguished due to gas alarms (State %u) |
N/A |
The plasma was extinguished due to an alarm relating to gas flows. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
The gas supply is not turned on. |
Turn on the gas supply. |
||
The gas supply pressure is not within the required range. |
Adjust the gas pressures to the ranges indicated in the ICP-OES Site Prep Guide or on the right side of the instrument. |
||
Plasma extinguished due to mains alarms (State %u) |
N/A |
The plasma was extinguished due to an alarm or lost communications with the Main Power Module. |
Restart the instrument from the front power button. Restart the instrument from the left side power switch. |
Plasma extinguished due to SSRF alarms (State %u) |
N/A |
The plasma was extinguished due to alarms reported by or communication lost with SSRF control PCA (board). The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Plasma extinguished due to open hardware interlocks (State %u) |
Plasma turned off. LED is red. |
The plasma was extinguished due hardware interlocks being triggered. The interlock switches did not actuate correctly. The plasma compartment door is not fully closed. |
Fully open the plasma compartment door and then close. |
Torch loader handle is not fully back in closed position. |
Fully open the torch loader handle and then close. |
||
Plasma extinguished due to water flow issue (State %u) |
N/A |
The plasma was extinguished due to an error with the water flow. |
Refer to the Error log for specific water error messages for troubleshooting advice. |
Fault returned for gas channel state (State %u) |
N/A |
The plasma was extinguished due to an error with gas flows. |
Refer to the Error log for specific gas error messages for troubleshooting advice. |
Timeout in plasma controller state operations (State %u) |
LED is red. |
During plasma ignition a condition required has not been met, and plasma ignition aborted. |
Retry plasma ignition. Refer to the Error log for other plasma errors. |
Received completion code %u to response to command %u |
N/A |
An error was returned from internal communications with P500 processes when attempting to perform plasma ignition. |
Retry plasma ignition. |
Failed to stabilize gas flow for plasma operations |
N/A |
A gas flow channel has reported an error during plasma stabilization operations. The gas flow has not been able to stabilize within the time limit. |
Refer to the Error log for specific gas error messages for troubleshooting advice. |
Communications error received while setting plasma channel gas flow (%.2f LPM) |
N/A |
An error was returned from internal communications with P500 gas controller process and/or gas control PCA when attempting to set plasma channel gas flow during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific gas error messages for troubleshooting advice. |
The instrument operation has become unstable |
Restart the instrument from the front power button. |
||
Communications error received while setting auxiliary channel gas flow (%.2f LPM) |
N/A |
An error was returned from internal communications with P500 gas controller process and/or gas control PCA when attempting to set auxiliary channel gas flow during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific gas error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting nebulizer channel gas flow (%.2f LPM) |
N/A |
An error was returned from internal communications with P500 gas controller process and/or gas control PCA when attempting to set nebulizer channel gas flow during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific gas error messages for troubleshooting advice.
|
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting oxygen channel gas flow (%.2f LPM) |
N/A |
An error was returned from internal communications with P500 gas controller process and/or gas control PCA when attempting to set oxygen injection channel gas flow during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific gas error messages for troubleshooting advice.
|
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting make-up channel gas flow (%.2f LPM) |
N/A |
An error was returned from internal communications with P500 gas controller process and/or gas control PCA when attempting to set make-up channel gas flow during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific gas error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting cone purge gas flow |
N/A |
"An error was returned from internal communications with P500 gas controller process and/or gas control PCA when attempting to set cone purge gas flow during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific gas error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting RF operation mode (Mode %u) |
N/A |
An error was returned from internal communications with P500 RF controller process and/or RF control PCA when attempting to set operation state during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific RF error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting RF target power |
N/A |
"An error was returned from internal communications with P500 RF controller process and/or RF control PCA when attempting to set target plasma power during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific RF error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting pump speed (%.1f RPM) |
N/A |
An error was returned from internal communications with P500 pump controller process and/or pump control PCA when attempting to set peristaltic pump speed during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific pump error messages for troubleshooting advice.
|
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting water cooling operations |
N/A |
An error was returned from internal communications with P500 chassis controller when attempting to enable RF water flow during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific water error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting mains blower power percentage (%.1f%%) |
N/A |
"An error was returned from internal communications with P500 mains controller process and/or mains control PCA when attempting to set blower speed during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific RF error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting RF operation mode (Mode %u) |
N/A |
An error was returned from internal communications with P500 RF controller process and/or RF control PCA when attempting to set target plasma power during plasma stabilization operations. |
Retry operation. Refer to the Error log for specific RF error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Communications error received while setting RF operation mode (Mode %u) |
N/A |
An error was returned from internal communications with P500 RF controller process and/or RF control PCA when attempting to set operation state during ignition operations. |
Retry operation. Refer to the Error log for specific RF error messages for troubleshooting advice. |
The instrument operation has become unstable |
Restart the instrument from the front power button. |
||
Gas channel flows not trending towards required set-points during ignition ramp-up |
Plasma ignition aborted LED is red. |
The instrument was unable to ramp gas flows correctly during ignition sequence. |
Verify gases are turned on and supply pressure set correctly. |
There is a problem with the torch. |
Check for any blockages and correct assembly. |
||
There is a problem with gas tubing for leaks, kinks or blockages. |
|||
There is a fault with the plasma gas module. |
Refer to the gas module troubleshooting advice. |
||
RF PCA failed to sustain plasma, oscillator exited running state |
N/A |
The RF System has reported an error. The RF control PCA (board) unexpectedly transitioned from running state. |
Retry operation. Refer to the Error log for specific RF error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Received completion code %u in response to command %u |
N/A |
Received error response from P500 gas controller when attempting to perform gas zeroing operation. |
Retry operation. Refer to the Error log for specific gas error messages for troubleshooting advice. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Lost SLIP communications with RF control PCA |
Plasma turned off. LED is red. |
The instrument controller unexpectedly lost communications with the RF Control PCA (board). |
Retry operation. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Error setting RF operation mode (Mode %u) |
Plasma turned off. LED is red. |
An error was returned when attempting to communicate with the RF Control PCA (board). |
Retry operation. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
Error setting RF coil target power (%.3f) |
Plasma turned off. LED is red. |
"An error was returned when attempting to set a plasma power. |
Retry operation. |
The instrument operation has become unstable. |
Restart the instrument from the front power button. |