If the problem persists after trying the following, contact your Agilent field service engineer or representative.
Error Message |
Instrument action and status LED |
Explanation |
Remedy |
Gas channel %u timed out with no response from PCA |
Plasma turned off. LED is red. |
After requesting a gas flow a time-out has occurred waiting for confirmation message. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Lost SLIP communications with gas control PCA |
Plasma is turned off. LED is red. |
The instrument controller has lost communications with the gas control module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Error setting gas channel flow/pressure (Channel %u, Level %.1f) |
Plasma turned off. LED is red. |
There has been a time-out or error returned from communications with Gas Control PCA (board) when attempting to set channel flow/pressure. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Error performing zero adjust operation |
Stop plasma ignition sequence. LED is red. |
There has been a time-out or error returned from communications with Gas Control PCA (board) when attempting to zero gas control module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Error setting polychromator boost purge flow (%.2f) |
LED is red. |
"There has been a time-out or error returned from communications with Gas Control PCA (board) when attempting to set boost purge. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Nebuliser channel has failed. Flow is outside tolerance. Target: %d mLPM, Actual %d mLPM |
Abort the run. Turn plasma off. |
The measured nebulizer flow has gone outside the set point. There is a problem with the nebulizer. |
|
There is an argon gas leak or blockage of the nebulizer argon fittings or tubing. |
|||
Makeup channel has failed. Flow is
outside tolerance. Target: %d mLPM, Actual %d mLPM |
Abort the run. Turn plasma off. |
The measured makeup flow has gone outside the set point. There is a problem with the device connected to the makeup gas connection. |
Consult the documentation that came with the device. |
There is an argon gas leak or blockage of the makeup fittings or tubing. |
|||
Auxillary channel has failed. Flow
is outside tolerance. Target: %d mLPM, Actual %d mLPM |
Abort the run. Turn plasma off. |
The measured plasma auxiliary flow has gone outside the set point. The torch is not loaded correctly to the torch loader. |
|
The torch is not assembled correctly. |
|||
There is a fault with the torch. |
|||
There is foreign material in the torch loader. |
Wipe down the torch loader. |
||
There is an argon gas leak or blockage of the plasma auxiliary fittings or tubing connected to the torch loader. |
|||
Plasma channel has failed. Flow is
outside tolerance. Target: %d mLPM, Actual %d mLPM |
Abort the run. Turn plasma off. |
The measured plasma flow has gone outside the set point. The torch is not loaded correctly to the torch loader. |
|
The torch is not assembled correctly. |
|||
There is a fault with the torch. |
|||
There is foreign material in the torch loader. |
Wipe down the torch loader. |
||
There is an argon gas leak or blockage of the plasma auxiliary fittings or tubing connected to the torch loader. |
|||
U13 Failed during Reset with error %d U14 Failed during Reset with error %d U23 Failed during reset with error %d The epc startup failed. Retrying. |
LED is red. |
There has been an error communicating from the purge gas control module to the plasma gas control module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
The atmospheric temperature is uncalibrated. P=%f |
Ignition not allowed. LED is red. |
An error occurred when attempting to read the atmospheric temperature calibration information from the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Calibration failed for Nebulizer
with error %d |
Ignition not allowed. LED is red. |
An error occurred when attempting to read the nebulizer flow calibration information from the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Calibration failed for MakeUp with
error %d |
Ignition not allowed. LED is red. |
An error occurred when attempting to read the makeup gas flow calibration information from the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Calibration failed for Auxiliary with error %d |
Ignition not allowed. LED is red. |
An error occurred when attempting to read the auxiliary plasma gas flow calibration information from the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Calibration failed for Plasma with error %d |
Ignition not allowed. LED is red. |
An error occurred when attempting to read the plasma gas flow calibration information from the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Calibration failed for write Nebulizer
with error %d |
Ignition not allowed. LED is red. |
An error occurred when attempting to write the makeup gas flow calibration information for the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Calibration failed for write Make Up with error %d |
Ignition not allowed. LED is red. |
An error occurred when attempting to write the nebulizer gas flow calibration information for the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
The setpoint for the flow is too
low. Channel: %u, Flow: %f |
Ignition not allowed. LED is red. |
The value set for the gas flow is too low for the plasma gas module. |
|
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
The setpoint for the flow is too high. Channel: %u, Flow: %f |
Ignition not allowed. LED is red. |
The value set for the gas flow is too high for the plasma gas module. |
|
The instrument operation has become unstable. |
Restart the instrument from the front power button. |
||
The I2C Driver returned an error
of %u while writing %u and is retrying |
Ignition not allowed. LED is yellow. |
A communication error occurred when attempting to write to the plasma gas module and is retrying communications. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
The I2C Driver returned an error of %u while writing %u The I2C Driver returned an error of %u while reading %u and is retrying The I2C Driver returned an error of %u while reading |
Ignition not allowed. LED is red. |
There is a communication error with the plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
The Argon instrument supply pressure is low |
Allow ignition attempt. Only fail if flow rate not achieved during ignition process. LED is yellow. |
The instrument argon gas supply pressure is too low. The instrument argon supply is turned off. |
Turn on the instrument argon supply. |
The laboratory argon supply pressure regulator is set too low. |
Increase the argon supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
||
The argon supply tubing or fittings are leaking, blocked or kinked. |
Check tubing for leaks and damage. Replace as needed. |
||
The argon inlet filter is blocked. |
|||
The Argon instrument supply pressure is high |
Allow ignition attempt. Only fail if flow rate not achieved during ignition process. LED is yellow. |
The instrument argon gas supply pressure is too high. The laboratory argon supply pressure regulator is set too high. |
Reduce the argon supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
The Polychromator purge supply pressure is low |
LED is yellow. |
The polychromator purge supply pressure is low. The purge flow to the polychromator may be incorrect as a result. The instrument polychromator purge supply is turned off. |
Turn on the instrument polychromator purge supply.
|
The laboratory polychromator purge supply pressure regulator is set too low. |
Increase the polychromator purge supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
||
The Polychromator purge supply pressure is high |
N/A |
"The polychromator purge supply pressure is too high. The purge flow to the polychromator may be incorrect as a result. The laboratory polychromator purge supply pressure regulator is set too high. |
Decrease the polychromator purge supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
The Ar/Oxygen gas mix supply pressure is low |
N/A |
The instrument argon/oxygen gas supply pressure is too low. The flow of argon/oxygen gas may be incorrect. The instrument argon/oxygen supply is turned off. |
Turn on the instrument argon/oxygen gas supply. |
The laboratory argon/oxygen gas supply pressure regulator is set too low. |
Increase the argon/oxygen gas supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
||
The argon/oxygen supply tubing or fittings are leaking, blocked or kinked. |
Check tubing for leaks and damage. Replace as needed. |
||
The argon/oxygen inlet filter is blocked. |
|||
The Ar/Oxygen gas mix supply pressure is high |
N/A |
"The instrument argon/oxygen gas supply pressure is too high. The flow of argon/oxygen gas may be incorrect. The laboratory argon/oxygen gas supply pressure regulator is set too high. |
Decrease the argon/oxygen gas supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
Cone flow high. Expecting %f, is %f |
Abort the run. Turn plasma off. LED is red. |
The cone purge flow is high. The cone purge pressure (flow) reached 5% of the argon inlet pressure. There is a blockage of the cone. |
|
Cone flow low. Expecting %f, is %f |
Abort the run. Turn plasma off. LED is red. |
The cone purge flow is low. There is a blockage of the cone. |
|
Snout flow low. Expecting %f, is %f |
LED is red. |
The snout purge flow is low. There is a blockage of the snout. |
|
Polychromator flow fault. Expecting %f, is %f |
N/A |
The polychromator purge flow is low. The instrument polychromator purge supply is turned off. |
Turn on the instrument polychromator purge gas supply. |
The laboratory polychromator purge supply pressure regulator is set too low. |
Increase the polychromator purge gas supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
||
The polychromator purge gas filter is blocked or leaking. |
|||
The polychromator purge tubing is blocked, kinked or damaged. |
Check tubing for leaks and damage. Replace as needed. |
||
Oxygen flow fault. Expecting %f, is %f |
N/A |
The argon/oxygen flow is outside of the set point permissible range. The instrument argon/oxygen supply is turned off. |
Turn on the instrument argon/oxygen gas supply. |
The laboratory argon/oxygen supply pressure regulator is set too low. |
Increase the argon/oxygen gas supply pressure on the pressure regulator. Consult the ICP-OES Site Prep Guide for required pressures. |
||
The argon/oxygen tubing is blocked, kinked or damaged. |
Check tubing for leaks and damage. Replace as needed. |
||
Previous transaction is not complete |
N/A |
"There has been a collision of requests to plasma gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
The manifold detection was inconclusive. ID pulse count = %d |
Enable oxygen addition. |
The inlet gas manifold detection is inconclusive. The gas manifold type is determined optically. The gas inlet manifold is not correctly installed. |
|
The gas inlet manifold is dirty or corroded. |
|||
Chain attempted to start before last
conversion was complete. |
N/A |
The purge gas module has become unresponsive. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Valve driver Fail pin has been pulled low for Channel %u |
N/A |
The purge gas module valve driver has reported an error. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
+15 V supply voltage under threshold +15 V supply voltage over threshold -15 V supply voltage under threshold -15 V supply voltage over threshold |
N/A |
The purge gas module has reported an error with a power supply voltage generated within the purge gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
+24 V supply voltage under threshold +24 V supply voltage over threshold |
N/A |
The purge gas module has reported an error with the 24V supply from the mains power module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
+3.3 V supply voltage under threshold +3.3 V supply voltage over threshold Gas control voltage reference under threshold Gas control voltage reference over threshold EPC controller voltage reference under threshold EPC controller voltage reference over threshold |
N/A |
The purge gas module has reported an error with a power supply voltage generated within the purge gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Revision ID (%u), is greater than firmware can handle (%u) |
N/A |
A hardware/firmware mismatch has been detected with the purge gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |
Error queue is full Checksum failed with length %u and checksum %u SLIP Manager timeout waiting to received a packet The watchdog caused the processor to reset (%d time) |
|
There is an error with the gas module. The instrument operation has become unstable. |
Restart the instrument from the front power button. |