Running the Instrument Performance Test

This test completes an analytical performance suite of tests including Resolution, Sensitivity (SRBR - Signal to Root Background Ratio), and Precision by running the multi-element standard and blank, then running the standard as a sample. The tests are run from the 'Tests' tab in the 'Instrument' window.

The data obtained from the tests is used to calculate SRBR and also measures short term precision and calibration accuracy.

If 'Advanced Valve System test' and 'Use autosampler' are selected on the Tests tab, the Instrument Performance test is not available.

To run both tests, deselect 'Use autosampler' and introduce the required solutions manually.

Before Starting the Instrument Performance Test

Running the Instrument Performance Test

  1. Click Instrument on the ICP Expert toolbar.
  2. Click the Tests tab.
  3. Deselect all tests except Instrument Performance.

    If you choose to use the autosampler instead of introducing samples manually, select Use Autosampler and then place the wavelength calibration, rinse, and blank solutions in the appropriate locations in the autosampler. Check the Autosampler tab on the worksheet page for solution locations.

  4. Click Run Tests.
  5. The method parameters are automatically filled in when the test begins.
  6. Follow the on-screen prompts to introduce solutions if you are manually introducing the solutions.

To print the report, click the print button in the top left corner of the print preview display.