There are a multitude of potential causes for an instrument exhibiting low sensitivity (low analytical signal intensity). These include, but are not limited to:
The first step is to run an 'Instrument Performance' test from the 'Tests' tab in the ICP Expert software to validate the instrument. This will perform the Resolution, Sensitivity, and Precision tests. Compare results to typical values.
Poor Sensitivity results will verify a problem exists. Using the generated test report, check which wavelengths are low on sensitivity and by how much. Compare to the installation results or past installation qualification numbers. The table below will give some direction as to failure mode and possible solution.
Instructions on how to run the 'Instrument Performance' test are here.
Failure Mode (Intensity) |
Possible cause |
Possible cause (Next level) |
All Wavelengths very low on sensitivity |
Sample introduction major blockage |
|
Leaking fitting
|
||
Sample delivery
|
||
Standard concentration |
Prepare new standard solutions |
|
Optimization of parameters |
||
Adjust the viewing height on the Conditions page for radial or SVDV instruments |
||
Adjust the Plasma flow or RF power on the Conditions page |
||
Wavelengths off peak
|
Instrument warm up - allow 20 minutes for the instrument to warm up from an idle state |
|
Dark current values high |
||
Contaminated blank |
Sample preparation - prepare a new blank solution |
|
Sample matrix |
||
Random Wavelengths low on Sensitivity |
Wavelengths off peak |
Instrument warm up - allow 20 minutes for the instrument to warm up from an idle state |
Peaks missing or low due to moisture forming on Detector Lens |
Moisture in purge lines. Refer to Contaminated Purge Gas. |
|
Dark current values high |
||
Short Wavelengths low on sensitivity |
Sample Introduction minor restriction |
|
Air present in optics (lack of purging) |
Optics purge - turned on automatically when the instrument is turned on. Check gas supply and lines for leaks or damaged tubing. |
|
Boost purge - provides a higher gas flow to the optics |
||
Cone purge - turned on automatically when the plasma is turned on.
|
||
Window Contamination |
||
Optimization of parameters |
||
Adjust the viewing height on the Conditions page for radial or SVDV instruments |
||
Adjust the Plasma flow or RF power on the Conditions page |
||
All Wavelengths marginal Sensitivity |
Wavelengths off peak |
Instrument warm up - allow 20 minutes for the instrument to warm up from an idle state |
Dark current values high |
Optics Purge gas contaminated with moisture is a potential cause of loss of sensitivity and in a number of cases random loss of peaks. Moisture from purge gas will collect on the CCD window resulting in loss of peaks for specific wavelengths.
To verify if this is a problem turn off the chiller, wait approximately 30 minutes and then switch chiller back on and run the Instrument Calibration. The moisture should have evaporated as the camera heats up and the calibration should again pass.
If this resolves the problem purge the gas lines and/or add a moisture filter to the purge lines.
Check the nebulizer for a blockage and clean if necessary.
Check for salt or carbon build-up (with organic work) on the injector tube. Clean the torch if necessary.
Pump tubing that has lost its elasticity will not deliver the sample to the nebulizer correctly. When the pump tubing appears limp or flattened, or does not require stretching to fit over the pump rollers, it should be replaced with new pump tubing.
Lower the nebulizer flow, lower the pump rate and/or use a smaller ID pump tube.
Prepare a fresh set of standards at the correct concentrations.
Develop and run a method for optimization and check that:
Optimize the operating conditions, including the nebulizer flow and torch viewing height (for radial only), until the optimum signal is obtained.
A number of things may result in a scan not being displayed. The following procedure may help identify the problem.
Perform a Read spectrum (using 1 ppm Mn solution using the 257.610 nm line) and check that a peak is visible within the scan window.
If no peak is displayed try performing a Read spectrum on another element.
- Aspirate a yttrium solution and check the plasma for the characteristic red/blue/red emission; or
- Repeat the scan (Read spectrum) for the element of interest. If a peak is still not displayed in the scan window aspirate a sodium sample and view the plasma. When the sodium yellow coloration appears in the plasma, perform a scan at the sodium 589.592 nm line. If a peak is still not observed, contact your local Agilent field service engineer.
Very low sample or standard concentrations may not produce an analytical signal. Increase the concentration of the analyte 10-fold or 100-fold.
Perform a Read spectrum (using 1 ppm Mn solution) and check that:
Optimize the operating conditions until the optimum signal is obtained.
Check the nebulizer for blockage and clean if necessary.