How to Create a Custom Rack
Use the Custom Rack Editor to create new autosampler rack configurations.
Click
in the Tools field on the Autosampler
page to open the Custom Rack Editor window.
About custom racks
- Custom rack definitions (the .txt file that contains all of the
rack settings) are also saved to a default folder in My Documents>
Agilent > ICP Expert > Autosampler and can be copied to other
computers with ICP Expert. The racks are then available to be used
in new worksheets.
- The numbers on the tubes in the custom racks window represent the
order in which the solutions in the tubes will be measured.
- Accessing the custom rack editor from the toolbar does not save
the custom rack in a worksheet.
- Accessing the custom rack editor from the worksheet Autosampler
page will save the new rack information with the worksheet once the
worksheet is saved.
Creating
a custom rack
To create a custom rack:
- Obtain the specifications of the rack you wish to use.
- Carefully measure any missing specifications. See Custom
rack editor fields for measurement examples.
- Enter the required information
into the custom rack editor fields mentioned below.
- Test the positioning of the
probe over multiple points on the rack to ensure the probe stops in
the center of the tubes.
- Determine the optimum probe
depth.
Custom
rack editor fields
Click
in the Tools field on the Autosampler
page to open the Custom Rack Editor window.
The File menu includes the following:
- Load - load previously created custom rack definitions
- Save as - save newly created custom rack definitions. The default
location is C:\Users\User\Documents\Agilent\ICP Expert\Autosamplers.
- Copy to clipboard - copies the rack definition information which
can then be pasted into applications such as Microsoft Notepad, Word,
or Excel.
The main fields are:
- Rack label - the visible, selectable rack name found on
the Autosampler page.
- Width (mm) - width of the custom rack
- Length (mm) - length of the custom rack
- Number of columns - the number columns on the custom rack
- Number of rows - the number of rows on the custom rack
- Tube diameter (mm) - the diameter of the tube used in the
custom rack
- Tube volume (mL) - the volume of the tube used in the custom
rack
- Tube row offset (mm) - the positioning of the left column
of tubes relative to the left side of the rack in the diagram. A value
of '0' represents the left edge of the rack passing through the center
of the tubes in the left column
For example:

- Tube column offset (mm) - the positioning of the top row
of tubes relative to the upper side of the rack in the diagram. A
value of '0' represents the top edge of the rack passing through the
center of the tubes in the top row
For example:

- Tube row separation (mm) - the distance between the center
line of each column of tubes
For example:

- Tube column separation (mm) - the distance between the center
line of each row of tubes
For example:

- Alternate row offset (mm) - the change in the positioning
of tubes in row 1, and every second row after that, along the X-axis
of the rack
For example:

- Alternate column offset (mm) - the change in the positioning
of tubes in column 1, and every second column after that, along the
Y-axis of the rack
For example:

- Row/Column order - changing the selection between these
two buttons will alter whether tube numbers are displayed (and used
by the autosampler) across columns or rows of the rack
Testing
and adjusting the probe position
To test the probe position:
- Right-click the rack position and then select the correct rack
from Rack type menu.
- Enter the appropriate probe depth in the Probe Control section
on the Autosampler page.
100 mm is a good starting point for most racks, however, if you
have tall racks you may want to start with a smaller number.
It is best practice to start with an empty rack, and to enter a
probe depth to ensure that the probe is higher than the top of the
rack (to avoid collisions), but close enough to visually inspect that
the position of the probe is in the center of the rack tube position.
- Double-click a tube position on the autosampler image on the Autosampler page to move the probe
to that position and confirm that the position of the probe is in
the center of the rack tube position.
- Adjust the settings in the custom rack editor if needed and repeat
Step 3.
- Repeat Step 3 for each corner of the rack.
- Insert tubes into the rack and then repeat Step 4 to confirm the
positioning.
Testing
and determining the probe depth
To test the probe depth:
- Enter the appropriate probe depth in the Probe Control section
on the Autosampler page.
"0" is equivalent to the probe not being lowered. The
default value of 155 mm is a good starting point and works for most
standard racks.
If your rack places the tube bottom higher than most standard racks,
adjust the probe depth accordingly.
- With a tube in the rack, use the Move and Lower Probe button on
the Autosampler page to help
determine the best probe depth and to check that the probe does not
hit the bottom of the tube.
Click here for information on other
types of standard racks.
See also: