Set Default Preferences for All Worksheets

The Preferences screen of the Options window enables you to specify default settings for new worksheets. It consists of these screens:

Select the instrument to display when working in Demo mode

Click File > Options > Preferences to access the page.

Model when instrument is not connected

Choose between the 5100/5110 or 5800/5900 instrument model when creating a new worksheet while the instrument is not connected.

Mode when instrument is not connected

Select the default instrument mode when creating a new worksheet while the instrument is not connected.

  • RV ICP-OES - allows only the options and behavior expected when a radial only (RV) instrument is connected.
  • VDV ICP-OES - allows only the options and behavior expected when a Dual View (VDV) instrument is connected.
  • SVDV ICP-OES - allows the options and behavior expected when a Synchronous Dual View (SVDV) instrument is connected.

Use instrument settings when connected

Select to use the instrument mode that is currently enabled on the ICP-OES, not the one selected in the software.

Enable prompts during analysis

Click File > Options > Preferences to access the page.

Specify default settings relating to running an analysis in the Run section.

Enable prompting for manual sampling

Select to display a message requesting the next sample when manually introducing samples.

Enable prompting on reuse of autosampler racks

Select to display a message when the autosampler rack requires changing during use.

Enable autosampler rack configuration confirmation on batch analysis

Select to display a message prompting the user to confirm the rack configuration in between each worksheet run when running a batch analysis.

Specify default background correction settings for all worksheets

Click File > Options > Preferences to access the page.

Background correction

Set the default type of background correction to use for each wavelength. Click the drop-down menu to make your selection.

Tell me more about background correction.

Specify default common and measurement conditions for all worksheets

Click File > Options > Preferences to access the page.

Specify default settings relating to common and measurement conditions.

Switch delay (for SVS 1, SVS 1+, and AVS 4)

Enter the required delay time (the time taken for the sample solution to pass through the tubing from the autosampler to the switching valve), in seconds. You will need to calculate this delay. Generally, it will equal the sample uptake delay minus 2 or 3 seconds depending on the tubing length between the switching valve and nebulizer. How do I calculate the switch delay?

Click here for AVS 6/7 information.

Stabilization time

Set the amount of time for the plasma to equilibrate before it collects data for each solution. Type the number, or use the up and down arrows.

Uptake time

Set the amount of time for the sample to reach the plasma. Type the number, or use the up and down arrows. How do I calculate the uptake time?

Use fast pump (for Uptake time)

Select to make the peristaltic pump run at high speed (currently 80 rpm) during sample uptake.

Rinse time

Set the amount of time for the rinse solution to wash the spray chamber. Type the number, or use the up and down arrows.

Use fast pump (for Rinse time)

Select to make the peristaltic pump run at high speed (currently 80 rpm) during rinsing.

Use IsoMist

Select to enable the IsoMist accessory for every new worksheet.

IsoMist temperature set point

Set the default IsoMist temperature to be used in every new worksheet.

IntelliQuant mode

Select between the Standard and Snapshot mode when collecting IntelliQuant data.

What is the difference between Standard and Snapshot mode?

Sample introduction

Set the way in which you would like to introduce your samples. Click the down arrow and choose 'Manual' to manually introduce samples, or 'Autosampler' to use an autosampler to introduce samples.

Rinse port position

Available only with the SPS 4 Autosampler with dual rinse mode enabled. Select the default rinse port position to be used for all new worksheets.

Pump speed

Set the flow rate of solution delivered to the nebulizer. Type the number, or use the up and down arrows.

Read time

Set the amount of time for which each measurement replicate will be read. Click the down arrow to make your selection.

Replicates

Set the number of times to repeat each read per solution. Type the number, or use the up and down arrows.

Enable custom replicates

Select to enable the use of custom replicates. Configure the number of replicates for blanks, standards, QC solutions and samples.

This option is only available in the ICP Expert Pro feature pack.

Viewing mode

Select from Radial, Axial or SVDV to choose the default viewing mode.

Specify default standards settings for all worksheets

Click File > Options > Preferences to access the page.

Specify default settings relating to standards.

Number of standards

Set the number of standards to run for each calibration. Type the number, or use the up and down arrows.

Correlation coefficient limit

Set the correlation coefficient limit, below which, the analysis will fail. The default value is 0.9990. Type the number, or use the up and down arrows.

Include a blank in calibration

Select to include a blank during calibration as a default selection.

Units

Set the default units that will be used throughout the ICP Expert software. Click the down arrow to make your selection.

Use standard additions

Select to have standard additions used in every new worksheet.

Use reagent blank

Select to collect a reagent blank during calibration. This option is only used when running in Standard Addition mode.

%RSE limit

Enter the default maximum % relative standard error limit used in every new worksheet. Calibration values that fall above this limit will fail.

Reactive dilution trigger

Enter the percentage of the highest calibration standard for each wavelength that will trigger a reactive dilution. The default value is 110%. For example, if the highest concentration of a calibration standard is 100 ppm, samples containing that element at concentrations of 110 ppm or greater will trigger a reactive dilution.

Calibration fit

Set the calibration fit. Click the down arrow to make your selection. Choose from Linear, Rational or Quadratic.

Tell me more about Calibration fit.

Weighted fit

Select to perform a weighted fit on results.

Tell me more about Weighted fit.

Force through None

Select to make the calibration graph pass use the lowest concentration as the origin.

Force through blank

Select to make the calibration graph pass through the blank.

Tell me more about Through blank.

Force through Origin

Select to make the calibration pass through 0, 0.

Calibration error

Set the acceptable calibration error. Type the number, or use the up and down arrows.

Reslope error

Set the acceptable reslope error. Type the number, or use the up and down arrows.

Specify default sequence settings for all worksheets

Click File > Options > Preferences to access the page.

Specify default sequence settings.

Advanced dilution system 2 prime

Select to require the Advanced dilution system 2 (ADS 2) to prime at the start of a run for all worksheets when an ADS 2 is enabled. 

Prime cycles

Enter the default number of times the ADS 2 primes before the start of a run.

End of run actions

Set what you would like the plasma and peristaltic pump to do at the end of a run. Click the down arrows to make your selection.

Set pump speed

Select to specify how fast you would like the pump to operate at the end of a run. Set a low pump speed to conserve rinse solution.

Turn plasma and pump off

Select to make the plasma and peristaltic pump turn off at the end of a run.

Turn plasma off and reduce pump speed

Select to make the plasma turn off at the end of a run and to have the peristaltic pump slow down. This helps to prevent tubing wear. Solution will not be consumed in this mode if an autosampler is being used because turning off of the plasma will put the probe in the park position.

Park autosampler

Select to move the autosampler probe to the Park position at the end of the run.

Rinse system at end of run

Select to rinse the system for a set amount of time before turning the system off at the end of a run.

Enable recalibration

Select to enable recalibration.

Number of samples

Set the number of samples. Type the number, or use the up and down arrows.

Weight

Set the weight. Type the number, or use the up and down arrows.

Volume

Set the volume. Type the number, or use the up and down arrows.

Dilution

Set the dilution. Type the number, or use the up and down arrows.

Pump speed

Set how fast you would like the peristaltic pump to operate. Type the number, or use the up and down arrows.

Rinse system

Set the amount of time for rinsing the system. Type the number, or use the up and down arrows.

End of run rinse port

Available only with the SPS 4 Autosampler with dual rinse mode enabled. Select the default rinse port position to be used at the end of a run all new worksheets.

Calibration every (samples)

Set the number of samples to analyze before performing a calibration. Type the number, or use the up and down arrows.

Reslope rate

Set the number of samples to analyze before performing a recalibration. Type the number, or use the up and down arrows.

Disable polychromator boost purge after run

Select to turn off polychromator boost purge once the run has finished.

Disable snout purge after run

Select to turn off the snout purge once the run has finished.

Switch chiller off after run

Select to turn off the chiller once the run has finished.

Specify default analysis settings for all worksheets

Specify default settings relating to the Analysis page of the worksheet.

Sort results

Set the order in which to display results. Choose from Alphabetical, Wavelength or Type.

Disable Delete Results

Prevents results from getting deleted.

Specify default QC settings for all worksheets

Lower limit

Set the standards/sample %recovery lower limit. Type the number, or use the up and down arrows.

Upper limit

Set the standards/sample %recovery upper limit. Type the number, or use the up and down arrows.

Difference

Set the maximum acceptable duplicate or serial dilution difference between the current and previous sample. Type the number, or use the up and down arrows.

Concentration limit

Set the defined element recovery test concentration limit. Type the number, or use the up and down arrows.

Error action

Set the default action when a QC error occurs. Click the down arrow to make your selection.

Stop

Select to stop the run when a QC test failure occurs.

Flag and continue

Choose this to make a note of a QC failure in the results, but continue with the run.

Recalibrate and repeat with samples

Choose this to perform a new calibration when a QC failure occurs, and repeat the QC test with samples run between the Calibration and the QC Test.

Recalibrate and repeat

Choose this to perform a new calibration when a QC failure occurs, and repeat the QC test.

Repeat and flag

If a QC fails with this error action applied, the QC analysis will be repeated immediately after the failed QC. If it fails a second time, the solution will be flagged but the run will continue. This can be applied to manually inserted QCs and rate QC blocks. Sample-paired QCs will both be repeated if one of them fails. The entire rate QC block that includes the failed QC solution will be repeated. Repeated solutions will have "Repeat" added to the solution label.

Repeat and stop

If a QC fails with this error action applied, the QC analysis will be repeated. If it fails a second time, the run will stop. This can be applied to manually inserted QCs and rate QC blocks. Sample-paired QCs will both be repeated if one of them fails. The entire rate QC block that includes the failed QC solution will be repeated. Repeated solutions will have "Repeat" added to the solution label.

Enable IntelliQuant and Enable QC

Select to include these options on all new worksheets. IntelliQuant is on by default.

Specify default settings relating to the autosampler racks for all worksheets

Standards rack

Select the rack size you want to use as the default standard rack size.

Rack

Select the rack size you want to use as the default sample rack size.

Probe speed

Select between normal (default) and viscous (slower) solutions to adjust the probe movement speed. This affects the following autosamplers:

  • SPS 3 and SPS 4 - up, down and rotational probe movement
  • ASX-520 - probe up movement only (when ICP Expert Pro is enabled)

ICP Expert Pro has a larger list of autosamplers available.

Rinse pump speed

Select the default rinse pump speed for the SPS 4 autosampler.

Reactive dilution aim

Choose the default target on the calibration curve that ICP Expert aims for when performing reactive dilutions. This option will be selected for all new worksheets when the ADS 2 or ESI PrepFAST autosampler accessory is selected on the Configuration page. 

Tell me more about reactive dilution.

Specify default Intelligent Rinse settings for all worksheets

Click here for detailed information on enabling, setting up and using Intelligent Rinse.

Enable Intelligent Rinse

Select to enable Intelligent Rinse on all new worksheets.

Washout rinse method

Select the default washout rinse method to be used on all new worksheets.

Reacquire baseline on timeout

Select to acquire a new baseline measurement if the maximum default rinse time limit is reached before washout signal has reached the defined limit.

Reset

Resets all changed values back to default.

See also: